Scanning Kelvin Probe System

The Scanning Kelvin Probe (SKP) is a non-contact, non-destructive instrument designed to measure the surface work function difference between conducting, coated, or semi-conducting materials and a metallic probe. The technique operates using a vibrating capacitance probe, and through a swept backing potential, the work function difference is measured between the scanning probe reference tip and sample surface. The work function can be directly correlated to the surface condition. A unique aspect of the SKP is its ability to make measurements in a humid or gaseous environment.  Applications include: -

 
scanning kelvin probe 
Figure 1
  • Filiform corrosion studies
  • Transport process of hydrated ions
  • Study of fuel cell catalysts
  • Forensic detection of fingerprints
  • Charaterisation of photovoltaic materials
  • Femi-level mapping

The principle of operation of the Kelvin Probe is illustrated in Figure 1.

The figure represents three energy level diagrams depicting two metals having a constant separation, work functions and Fermi levels of (φ1, E1) and (φ2, E2), respectively.

The first diagram shows the metals with no electrical contact and differing Fermi levels. When electrical contact is made (middle diagram) the flow of charge allows the Fermi levels to equalize and gives rise to a surface charge.

 

The potential difference, VC, is related to the difference in work function:

-eVC = φ1 - φ2 (1) where e is the electron charge.

The inclusion of an external backing potential, VB (final diagram), allows for a nulling of the surface charge at a unique point where VB = - VC. This point represents the work function difference between the two materials.  Work function can then be used to determine the corrosion potential (Ecorr) at a specific point using the relationship:

Ecorr = Constant + (φ1 - φ2) (2) where (φ1 - φ2) is the measured work function between the probe and the sample. 

 
filiform corrosion 
Figure 2

The constant can be determined by measuring the corrosion potential using a conventional reference electrode in an electrolyte. Once the constant is known for a particular sample, Ecorr can be calculated directly from the SKP data without an electrolyte.

 

Figure 2 shows the Scanning Kelvin Probe Measurement of filiform corrosion on Aluminium

 
 

Figure 3

Figure 3 shows a surface map of the Scanning Kelvin Probe signal recorded over the heat affected zone of welded material.

 
 

Figure 4


 

Figure 4 shows a surface map of the Scanning Kelvin Probe signal recorded over a printed circuit board.

如果您對本品感興趣請與我公司
北京東方安諾生化科技有限公司
:    -8007


: www.annovatech.com
: caoye513



化工儀器網(wǎng)

采購商登錄
記住賬號    找回密碼
沒有賬號?免費注冊

提示

×

*您想獲取產(chǎn)品的資料:

以上可多選,勾選其他,可自行輸入要求

個人信息:

溫馨提示

該企業(yè)已關(guān)閉在線交流功能

主站蜘蛛池模板: 安图县| 广宁县| 扶绥县| 蓬莱市| 招远市| 塔河县| 平泉县| 灵石县| 龙山县| 额济纳旗| 那曲县| 永宁县| 临沧市| 花莲县| 芜湖市| 仁寿县| 丰原市| 宣城市| 胶州市| 布拖县| 北碚区| 谷城县| 南召县| 河池市| 衢州市| 东丽区| 金湖县| 拉萨市| 开江县| 新郑市| 辛集市| 祁东县| 武宁县| 勐海县| 伊通| 孟州市| 集安市| 武威市| 霍州市| 霍州市| 苍溪县|